Standard Guide for Biocompatibility Evaluation of Medical Device Packaging Materials
Automaticky přeložený název:
Standardní Příručka pro biokompatibility hodnocení zdravotnického prostředku obalových materiálů
NORMA vydána dne 1.4.2005
Označení normy: ASTM F2475-05
Poznámka: NEPLATNÁ
Datum vydání normy: 1.4.2005
Kód zboží: NS-53957
Počet stran: 4
Přibližná hmotnost: 12 g (0.03 liber)
Země: Americká technická norma
Kategorie: Technické normy ASTM
Keywords:
biocompatibility, cytotoxicity, medical device, medical device packaging, toxicity, ICS Number Code 11.120.99 (Other standards related to pharmaceutics), 55.020 (Packaging and distibution of goods in general)
Significance and Use | ||
The compatibility of packaging materials with a medical device is a requirement of many regulatory bodies. Since most medical devices are used or implanted in, around or on the human body, these devices must do no harm. Therefore, the packaging materials that come in contact with the medical device must also be evaluated and determined to be safe for use with the human body in that they have no negative impact on the physical, chemical or biological properties of the device. . This evaluation may include both a study of relevant experience with and actual testing of packaging materials. Such an evaluation may result in the conclusion that no testing is needed if the material has a demonstrable history of safe use in the specific role that is the same as that of the package under design. The medical device manufacturer determines the need for appropriate testing, with consideration of the device/package interactions, if any. The responsibility of the packaging supplier is typically limited to the performance of cytotoxicity testing. |
||
1. Scope | ||
1.1 This guide provides information to determine the appropriate testing for biocompatibility of packaging materials used to contain a medical device. 1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and to determine the applicability of regulatory limitations prior to use. |
||
2. Referenced Documents | ||
|
Chcete mít jistotu, že používáte pouze platné technické normy?
Nabízíme Vám řešení, které Vám zajistí měsíční přehled o aktuálnosti norem, které používáte.
Chcete vědět více informací? Podívejte se na tuto stránku.
Poslední aktualizace: 12.05.2025 (Počet položek: 2 199 201)
© Copyright 2025 NORMSERVIS s.r.o.