Standard Test Method for Determination of Volatile Organic Compounds in Atmospheres (Canister Sampling Methodology)
Přeložit název
NORMA vydána dne 1.12.2015
Označení normy: ASTM D5466-15
Poznámka: NEPLATNÁ
Datum vydání normy: 1.12.2015
Kód zboží: NS-640484
Počet stran: 21
Přibližná hmotnost: 63 g (0.14 liber)
Země: Americká technická norma
Kategorie: Technické normy ASTM
Keywords:
ambient atmospheres, air analysis, canister analysis, canister sampling, gas chromatography-mass spectrometry, indoor atmospheres, volatile organic compounds, workplace atmospheres,, ICS Number Code 13.040.01 (Air quality in general)
Significance and Use | ||||||||||
5.1 VOCs are emitted into ambient, indoor, and workplace air from many different sources. These VOCs are of interest for a variety of reasons including participation in atmospheric chemistry and acute or chronic human health impacts. 5.2 Canisters are particularly well suited for the collection and analysis of very volatile and volatile organic compounds. This test method describes the collection and analysis of whole gas samples and is not subject to high volatility limitations. 5.3 Chemically stable selected VOCs have been successfully collected in passivated stainless steel canisters. Collection of atmospheric samples in canisters provides for: (5.4 Interior surfaces of the canisters may be treated by any of several proprietary passivation processes including an electropolishing process to remove or cover reactive metal sites on the interior surface of the vessel and a fused silica coating process. 5.5 For this standard, VOCs are defined as organic compounds that can be quantitatively recovered from the canisters having a vapor pressure greater than 10-2 kPa at 25ºC and 760 mm Hg. 5.6 Target compound polarity is also a factor in compound recovery. Aliphatic and aromatic hydrocarbons from C1 to C13 have been successfully measured with this standard but are not listed in Table 1 5.7 Recovery and stability studies shall be conducted on compounds not listed in Table 1 before expanding the use of this test method to additional compounds. |
||||||||||
1. Scope | ||||||||||
1.1 This standard describes a procedure for sampling and analysis of selected volatile organic compounds (VOCs) in ambient, indoor, and workplace atmospheres. The test method is based on the collection of whole air samples in stainless steel canisters with specially treated (passivated) interior surfaces. For sample analysis, a portion of the sample is subsequently removed from the canister and the collected VOCs are selectively concentrated by adsorption or condensation onto a trap, subsequently released by thermal desorption, separated by gas chromatography, and measured by a mass spectrometric detector or other detector(s). This test method describes procedures for sampling into canisters to final pressures both above and below atmospheric pressure (respectively referred to as pressurized and subatmospheric pressure sampling).2 1.2 This test method is applicable to specific VOCs that have been determined to be stable when stored in canisters. Numerous compounds, many of which are chlorinated VOCs, have been successfully tested for storage stability in pressurized canisters 1.3 The procedure for collecting the sample involves the use of inlet lines, air filters, flow rate regulators for obtaining time-integrated samples, and in the case of pressurized samples, an air pump. Typical long-term fixed location canister samplers have been designed to automatically start and stop the sample collection process using electronically actuated valves and timers 1.4 The organic compounds that have been successfully measured at single-digit parts-per-billion by volume (ppbv) levels with this test method are listed in Table 1. The test method is applicable to VOC concentrations ranging from the detection limit to approximately 300 ppbv. Above this concentration, smaller sample aliquots of sample gas may be analyzed or samples can be diluted with dry ultra-high-purity nitrogen or air. 1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Safety practices should be part of the user's SOP manual. |
||||||||||
2. Referenced Documents | ||||||||||
|