Standard Test Method for Determination of Calcium, Magnesium, Potassium, and Sodium in Atmospheric Wet Deposition by Flame Atomic Absorption Spectrophotometry
Automaticky přeložený název:
Standardní testovací metoda pro stanovení vápníku, hořčíku , draslíku a sodíku v atmosférických mokré depozice plamenovou atomovou absorpční spektrofotometrií
NORMA vydána dne 1.4.2008
Označení normy: ASTM D5086-01(2008)
Poznámka: NEPLATNÁ
Datum vydání normy: 1.4.2008
Kód zboží: NS-29780
Počet stran: 8
Přibližná hmotnost: 24 g (0.05 liber)
Země: Americká technická norma
Kategorie: Technické normy ASTM
Keywords:
atmospheric wet deposition, atomic absorption, calcium, magnesium, potassium, precipitation, sodium, Atmospheric wet deposition (AWD), Beer&apos,s law, Calcium content--atmospheric analysis, Contamination--atmospheric, FAAS (flame atomic absorption spectrometry), Hail, Magnesium content--atmospheric analysis, Method detection limit (MDL), Potassium content--atmospheric analysis, Rain, Sleet, Snow, Sodium content--atmospheric analysis, Spectrophotometry--atomic absorption
| Significance and Use | ||||||||||||||||||||||||
|
This test method may be used for the determination of calcium, magnesium, potassium, and sodium in atmospheric wet deposition samples. Emphasis is placed on the easily contaminated quality of atmospheric wet deposition samples due to the low concentration levels of dissolved metals commonly present. Annex A1 represents cumulative frequency percentile concentration plots of calcium, magnesium, potassium, and sodium obtained from analyses of over five thousand wet deposition samples. These data may be used as an aid in the selection of appropriate calibration standard concentrations. (3) |
||||||||||||||||||||||||
| 1. Scope | ||||||||||||||||||||||||
|
1.1 This test method is applicable to the determination of calcium, magnesium, potassium, and sodium in atmospheric wet deposition (rain, snow, sleet, and hail) by flame atomic absorption spectrophotometry (FAAS). (1) 1.2 The concentration ranges are listed below. The range tested was confirmed using the interlaboratory collaborative test (see Table 1 for a statistical summary of the collaborative test).
1.3 The method detection limit (MDL) is based on single operator precision (2) and may be higher or lower for other operators and laboratories. Many workers have found that this test method is reliable at lower levels than were tested, but the precision and bias data presented are insufficient to justify their use at lower levels. 1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Specific warning statements are given in 8.3, 8.7, 12.1.8, and Section 9. |
||||||||||||||||||||||||
| 2. Referenced Documents | ||||||||||||||||||||||||
|
Chcete pravidelně odebírat informace o nově vycházejících normách z celého světa a to zcela zdarma?
Přihlašte se k odběru. Vše je velice jednoduché a absolutně ZDARMA.
Na výběr máte vydavatele z celého světa.
Poslední aktualizace: 08.09.2026 (Počet položek: 2 300 134)
© Copyright 2026 NORMSERVIS s.r.o.