Standard Practice for Choosing Sorbents, Sampling Parameters and Thermal Desorption Analytical Conditions for Monitoring Volatile Organic Chemicals in Air (Includes all amendments And changes 3/3/2023).
NORMA vydána dne 1.11.2015
Označení normy: ASTM D6196-15e1
Poznámka: NEPLATNÁ
Datum vydání normy: 1.11.2015
Počet stran: 31
Přibližná hmotnost: 93 g (0.21 liber)
Země: Americká technická norma
Kategorie: Technické normy ASTM
Keywords:
Place the sorbent or carrier tube in a compatible thermal desorption apparatus. As each tube in turn is sealed into the analytical flow path, system integrity should be checked to ensure there are no leaks which could lead to sample losses. Purge the air from each tube before heat is applied to avoid chromatographic artifacts arising from oxidation of the sorbent or damage to the chromatographic system. Heat the tube to displace the organic vapors which are pas